Proximal humeral internal locking system (PHILOS) for the treatment of proximal humeral fractures
Adult
Aged, 80 and over
Male
Reoperation
Humeral Fractures
Bone Screws
Middle Aged
Radiography
Fracture Fixation, Internal
03 medical and health sciences
Postoperative Complications
Treatment Outcome
0302 clinical medicine
Humans
Female
Bone Plates
Fractures, Malunited
Aged
DOI:
10.1007/s00402-006-0256-9
Publication Date:
2006-12-21T10:31:20Z
AUTHORS (7)
ABSTRACT
The aim of this study was to describe early results of a new internal locking system, PHILOS, used for the treatment of proximal humeral fractures.A chart and radiographic review of 25 cases that had proximal humeral internal locking system (PHILOS) plate for the treatment of proximal humeral fractures was performed.Of the 25 cases, 20 went to union with a mean neck/shaft angle of 127.2 degrees . Five cases required or were considered for revision surgery for non-union or implant failure. Of the 25 implants, 4 had screw protrusion into the gleno-humeral joint, 4 had screw loosening and backing out, and 1 plate broke without further trauma.Our results suggest that PHILOS is an effective system for providing fracture stabilisation to bony union but awareness of potential hardware complications is essential.
SUPPLEMENTAL MATERIAL
Coming soon ....
REFERENCES (16)
CITATIONS (79)
EXTERNAL LINKS
PlumX Metrics
RECOMMENDATIONS
FAIR ASSESSMENT
Coming soon ....
JUPYTER LAB
Coming soon ....