Proximal humeral internal locking system (PHILOS) for the treatment of proximal humeral fractures

Adult Aged, 80 and over Male Reoperation Humeral Fractures Bone Screws Middle Aged Radiography Fracture Fixation, Internal 03 medical and health sciences Postoperative Complications Treatment Outcome 0302 clinical medicine Humans Female Bone Plates Fractures, Malunited Aged
DOI: 10.1007/s00402-006-0256-9 Publication Date: 2006-12-21T10:31:20Z
ABSTRACT
The aim of this study was to describe early results of a new internal locking system, PHILOS, used for the treatment of proximal humeral fractures.A chart and radiographic review of 25 cases that had proximal humeral internal locking system (PHILOS) plate for the treatment of proximal humeral fractures was performed.Of the 25 cases, 20 went to union with a mean neck/shaft angle of 127.2 degrees . Five cases required or were considered for revision surgery for non-union or implant failure. Of the 25 implants, 4 had screw protrusion into the gleno-humeral joint, 4 had screw loosening and backing out, and 1 plate broke without further trauma.Our results suggest that PHILOS is an effective system for providing fracture stabilisation to bony union but awareness of potential hardware complications is essential.