A 95-dB DR second order incremental ΣΔ ADC for multi-channel healthcare application
Successive approximation ADC
Analog-to-digital converter
DOI:
10.1007/s10470-014-0479-6
Publication Date:
2015-01-23T17:50:10Z
AUTHORS (5)
ABSTRACT
To meet the requirement of multi-channel healthcare applications, a second order incremental ΣΔ ADC in 0.35µm CMOS technology is proposed for minimizing power consumption and achieving high resolution. The ADC achieves 95dB dynamic range in 240Hz bandwidth. The total power consumption of the ADC is 80µW and the chip core area is 0.21mm2.