Time and cost effective production of sodium [18F]fluoride using a dedicated automation module with disposable cassettes for GMP environment
Positron emission tomography
03 medical and health sciences
0302 clinical medicine
Sodium [18F]fluoride
Bone scintigraphy
DOI:
10.1007/s10967-016-4709-5
Publication Date:
2016-02-22T05:55:35Z
AUTHORS (7)
ABSTRACT
The aim of the present study is to establish a time and cost effective Na[18F]F preparation method. The process is composed of absorbing radioactive impurities with cation exchange cartridge (CM), trapping 18F-fluroide with QMA cartridges, and formulating the product. The total preparation time determined at the end of bombardment was 4 min and the radiochemical yield was 82 % (n = 29). The final formulated Na[18F]F solution satisfies the USP criteria for quality control.