A Capillary Fraction Collector Coupled to a Fluorescence Reader: A Novel Device to Continuously Quantify Glutamate During Microdialysis
Microdialysis
Neurochemical
Dialysis tubing
DOI:
10.1007/s11064-012-0736-y
Publication Date:
2012-03-07T06:23:04Z
AUTHORS (5)
ABSTRACT
Microdialysis coupled to HPLC is the preferred method for quantification of glutamate (Glu) concentrations, both in normal and pathological conditions. However, HPLC is a time consuming technique that suffers from poor temporal resolution. Here we describe an alternative method to measure glutamate concentrations in small-volume dialysis samples by quantifying hydrogen peroxide released by glutamate oxidase using the Amplex Red method. This system permits continuous automatic sample collection and the detection of a fluorescent reaction product, resorufin, which provides a measure of the glutamate concentration. Quantification can be carried out in small microdialysis samples to allow a temporal resolution of 60 s. Both in vitro and in vivo tests showed that this method was reproducible and reliable, detecting Glu along a linear scale. To validate the proposed method, extracellular Glu concentrations in the rat brain were measured and correlated with electrophysiological activity prior, during and after seizure induction with 4-aminopyridine. This method may be adapted to monitor other biologically active compounds, including acetylcholine and glucose, as well as other compounds that generate hydrogen peroxide as a reaction product and may be used as an alternative to other neurochemical methods.
SUPPLEMENTAL MATERIAL
Coming soon ....
REFERENCES (19)
CITATIONS (10)
EXTERNAL LINKS
PlumX Metrics
RECOMMENDATIONS
FAIR ASSESSMENT
Coming soon ....
JUPYTER LAB
Coming soon ....