ABSTRACT
In this work, we developed a microfluidic paper-based analytical device (μPAD) for the on-line isolation of proteins and the electrochemical detection of lead ions (Pb(II)) in urine samples. The patterned filter paper was prepared through the direct printing of microchannel patterns on filter paper using an office laser printer. The paper was modified with protein precipitant and was then coupled with a detachable three-electrode system. Experimental parameters, namely, modification reagents, microchannel length and width, deposition potential, and deposition time, were optimized. Then, the maximum protein concentration under which the device can function was obtained as 300 mg L-1. The linear range was 10-500 μg L-1 with a detection limit of 9 μg L-1. The effectiveness of this device was demonstrated through the quantification of Pb(II) in urine samples and the results agreed with those of atomic absorption spectrometry (AAS).