Spatially resolved measurement of wideband prompt gamma-ray emission toward on-line monitor for the future proton therapy
Wideband
Line (geometry)
DOI:
10.1016/j.nima.2017.10.020
Publication Date:
2017-10-27T15:33:30Z
AUTHORS (8)
ABSTRACT
Abstract In proton therapy, the delivered dose should be monitored to a high degree of accuracy to avoid unnecessary exposure to healthy tissues and critical organs. Although positron emission tomography (PET) is most frequently used to verify the proton range, the nuclear reactions between protons and nuclei that generate positrons do not necessarily correspond to the actual proton range. Moreover, such imaging must be conducted after the treatment irradiation, because a PET gantry cannot be used in conjunction with a proton therapy beam. In this paper, we studied one-dimensional (1D) and two-dimensional (2D) distributions of prompt gamma rays of various energies, to determine the most suitable energy window for online monitoring in proton therapy. After an initial simulation study using the particle and heavy ion transport code system (PHITS), we irradiated a poly(methyl methacrylate) (PMMA) phantom with a 70-MeV proton beam to mimic proton range verification in a clinical situation. Using a newly developed Compton camera, we have experimentally confirmed for the first time that 4.4-MeV gamma rays emitted from 12 C and 16 O match the exact position of the Bragg peak in proton range verification.