Evaluation of the efficacy of a metal artifact reduction algorithm in different cone beam computed tomography scanning parameters

03 medical and health sciences 0302 clinical medicine Phantoms, Imaging Humans Radiographic Image Interpretation, Computer-Assisted Reproducibility of Results Cone-Beam Computed Tomography Artifacts Head Algorithms
DOI: 10.1016/j.oooo.2017.02.015 Publication Date: 2017-03-09T07:33:00Z
ABSTRACT
The aim of this study was to evaluate the efficacy of a metal artifact reduction (MAR) algorithm in cone beam computed tomography (CBCT) images of dental materials obtained with different field-of-view (FOV) and voxel sizes.Two imaging phantoms were custom-made of acrylic resin. Each phantom had 3 cylinders made of the same dental material: dental amalgam or copper-aluminum alloy. CBCT scans were obtained separately for each of the imaging phantoms using the Picasso-Trio CBCT (Vatech, Hwaseong, Republic of Korea) unit at 4 FOV sizes and 2 voxel sizes. Each imaging phantom was scanned with and without MAR. All images were evaluated in the OnDemand3D software (Cybermed, Seoul, Republic of Korea) and image noise (gray value variability) was calculated as the standard deviation (SD) of the gray values of regions of interest around the dental material cylinders. Data were compared by the Friedman test and Dunn test (α = 0.05). Intraclass correlation coefficient (ICC) was calculated to assess intraobserver reliability.MAR significantly reduced (P < .05) image noise around the dental materials, irrespective of FOV and voxel sizes, with an ICC of 0.997.The efficacy of MAR was similar for the different FOV and voxel sizes studied. Hence, imaging protocols and the use of MAR algorithm should be based on the selection criteria.
SUPPLEMENTAL MATERIAL
Coming soon ....
REFERENCES (19)
CITATIONS (33)