Design of amperometric urea biosensor based on self-assembled monolayer of cystamine/PAMAM-grafted MWCNT/Urease
Amperometry
Cystamine
Linear range
DOI:
10.1016/j.snb.2017.06.161
Publication Date:
2017-06-27T14:17:25Z
AUTHORS (3)
ABSTRACT
Abstract In this work, we report novel amperometric urea biosensor based on immobilization of Urease enzyme onto self-assembled polyamidoamine grafted multiwalled carbon nanotube (MWCNT-PAMAM) dendrimers (from G1, first generation up to G5) on Au electrode. Bio-sensing electrode’s performance parameters such as optimum pH, temperature, and applied potential together with reproducibility were investigated. The electrode based on fifth generation of MWCNT-PAMAM and urease enzyme showed excellent performance in urea detection at an applied potential of 0.45 V with short response time of 3 s, wide linear range of 1–20 mM, detection limit of 0.4 mM, and sensitivity of 6.6 nA/mM. No significant response was detected in the presence of possible interferon molecules. Non-diluted human blood plasma has been used for real sample application where electrode showed excellent performance in urea detection.
SUPPLEMENTAL MATERIAL
Coming soon ....
REFERENCES (67)
CITATIONS (89)
EXTERNAL LINKS
PlumX Metrics
RECOMMENDATIONS
FAIR ASSESSMENT
Coming soon ....
JUPYTER LAB
Coming soon ....