Evaluation of Dual-Frequency Switching HIFU for Optimizing Superficial Ablation

High-intensity focused ultrasound Focused Ultrasound Thermal ablation
DOI: 10.1016/j.ultrasmedbio.2024.02.016 Publication Date: 2024-03-27T13:59:46Z
ABSTRACT
Dual-frequency high-intensity focused ultrasound (HIFU) thermal ablation is an exceptionally promising technique for treating tumors due to its precision and effectiveness. However, there are still a few studies on improving the accuracy and efficiency of HIFU in superficial ablation applications. This study proposes a method utilizing dual frequency switching ultrasound (DFSU) to enhance the efficiency and precision of superficial treatments.A dual-frequency HIFU transducer operating at 4.5 MHz and 13.7 MHz was designed, and a dual-frequency impedance matching network was designed to optimize electro-acoustic conversion efficiency. Phantom and ex vivo tests were conducted to measure and compare thermal lesion areas and temperature rises caused by single-frequency ultrasound (SFU) and DFSU.In both phantom and ex vivo tests, the utilization of DFSU resulted in larger lesion areas compared to SFU. Moreover, DFSU provided improved control and versatility, enabling precise and efficient ablation.DFSU exhibits the ability to generate larger ablation areas in superficial tissue compared to SFU, and DFSU allows flexible control over the ablation area and temperature rise rate. The acoustic power deposition of HIFU can be optimized to achieve precise ablation.