Pulse Shape Analysis of a Segmented Planar Germanium Detector by Using the Green's Function Method and Rise-Time Correlations for Depth Information

Semiconductor detector Position (finance)
DOI: 10.1143/jpsj.73.3312 Publication Date: 2005-02-10T01:52:31Z
ABSTRACT
We performed pulse shape analysis to extract depth information for an interaction point of γ rays in a 25-fold segmented germanium detector. With 90° Compton spectrometer with collimators on both incidence and exit sides, we measured shapes the two points, ( x , y z ) = (2.5 cm, 2.5 0.5 cm) 1.0 cm). Experimental were compared calculated ones by three-dimensional Green's function method. Analytical solutions induced charges each cathode anode obtained hole electron as time. It is suggested that rise-time correlation between central signals, T 90 (anode) - 50 (cathode), can be used useful means determine given in-plane position ). shown correlations are qualitative agreement experimental uncollimated rays.
SUPPLEMENTAL MATERIAL
Coming soon ....
REFERENCES (34)