Computerized determination of pneumotachometer characteristics using a calibrated syringe

03 medical and health sciences 0302 clinical medicine Computers Calibration Pulmonary Ventilation Respiratory Function Tests
DOI: 10.1152/jappl.1982.53.1.280 Publication Date: 2017-12-24T23:58:48Z
ABSTRACT
A computerized method has been developed to determine the conductance characteristics of pneumotachometers. Conductance values flowmeter, which correspond all pressure values, can be determined by a weighted averaging technique, when multiple strokes precision 3-liter calibrated syringe are used. The then allow measured differential pressures converted, point point, into flows. accuracy volumes is within +/- 0.5% for 100-stroke calibration process, and with 50 provides 1% accuracy. improves on-line measurements ventilation. O2 consumption, CO2 production during exercise.
SUPPLEMENTAL MATERIAL
Coming soon ....
REFERENCES (0)
CITATIONS (73)