Rapid testing of red blood cell parameters in primary care patients using HemoScreen™ point of care instrument
Point-of-Care Testing
Point of care
White blood cell
Blood cell
Complete blood count
Blood count
Cuvette
DOI:
10.1186/s12875-019-0971-2
Publication Date:
2019-06-07T06:02:54Z
AUTHORS (4)
ABSTRACT
Patients with anemia are frequently encountered in primary care. Once is detected, it essential to define the type and identify underlying cause prior initiation of treatment. In most cases, can be determined using information from patient history, physical exam, complete blood counts (CBC). Point care testing cell would speed up work patients. The aim present study was evaluate if HemoScreen™ instrument (PixCell Medical, Yokneam Ilit, Israel) could used for samples. It a POCT that utilizes single sample cuvettes image analysis full count including RBC, Hemoglobin, MCV, MCH, platelets, WBC, WBC 5-part differential. We compared Sysmex XN results 100 samples focusing on total white cells, red parameters MCV platelets. Deming correlations between instruments CBC were WBCHemoScreen™ = 1.016* WBCSysmex + 0.34; r 0.981, RBCHemoScreen™ 0.988* RBCSysmex 0.015; 0.974, HemoglobinHemoScreen™ 1.081* HemoglobinSysmex − 11.25; 0.964, MCHHemoScreen™ 0.978* MCHSysmex 0.78; 0.939, MCVHemoScreen™ 0.963* MCVSysmex 8.68; 0.946, PlateletsHemoScreen™ 0.964* PlateletsSysmex 25.7; 0.953. provide rapid accurate test evaluation This new technology interesting as allows also at small centers.