Performance comparison of sub-miniature hotwire sensors used in a new Constant Temperature Anemometer
Anemometer
Constant (computer programming)
Temporal resolution
DOI:
10.1615/thmt-23.140
Publication Date:
2024-02-14T15:00:08Z
AUTHORS (2)
ABSTRACT
The spatial resolution of Constant Temperature Anemometry (CTA) measurements is limited by the length (l) sensor, due to averaging smaller turbulent eddies over sensor (spatial filtering). There are several limitations and challenges connecting dimensions measurements, which explained in detail Ref. [1] Section 5.2. One limitation l/d ratio, should be kept above 200 minimize end conduction effects. This indirectly puts a on diameter (d) means minimizing also reduction its diameter. As side benefit, achieving good reducing size temporal optimizing CTA bridge at design stage.