Microscopic Measurement of Circular Dichroism Spectra

Phaseolus Plant Leaves Circular Dichroism Membranes, Artificial DNA Equipment Design Microscopy, Polarization Particle Size Sensitivity and Specificity 01 natural sciences 0104 chemical sciences
DOI: 10.2116/analsci.24.297 Publication Date: 2008-02-12T05:49:43Z
ABSTRACT
A microscope device to measure the circular dichroism (CD) spectra of a specified microscopic region of chiral samples was constructed by combining of a couple of objective lenses and a CCD camera, which was installed in a sample chamber of a commercially available CD spectropolarimeter. By using this apparatus, high quality micro-CD spectra in the 60 x 60 microm region of samples could be measured. Micro-CD spectra of thin film of chiral DNA samples on glass and a natural kidney bean leaf were measured, and the potential of the micro-CD apparatus was successfully demonstrated.