Magnetic Solid-Phase Extraction in Combination with Ultra High-Performance Liquid Chromatography-Tandem Mass Spectrometry Analysis: Quantification of Risperidone and 9-Hydroxyrisperidone in Biological Samples

risperidone UPLC-MS/MS biological sample 9-hydroxyrisperidone magnetic solid-phase extraction 01 natural sciences 0104 chemical sciences
DOI: 10.21577/0103-5053.20170016 Publication Date: 2017-02-06T18:37:47Z
ABSTRACT
Oleic acid modified Fe3O4 nanoparticles were proposed for magnetic solid-phase extraction coupled with ultra high-performance liquid chromatography-tandem mass spectrometry to determine risperidone (RISP) and 9-hydroxyrisperidone (9OHR) in biological samples. The effects of various experimental parameters including adsorbent amount, pH, adsorption time, eluent type and concentration were systematically investigated. Under optimal conditions, the calibration curve was linear within the concentration range of 0.2-200 ng mL-1 with the correlation coefficient (r) of 0.9962 and the lower limit of quantification was 0.06 ng mL-1 for RISP; the calibration curve was linear within the concentration range of 0.2-200 ng mL-1 with the correlation coefficient of 0.9956 and the lower limit of quantification was 0.04 ng mL-1 for 9OHR. The extraction recoveries were over 90.0% with relative standard deviation less than 5.0%. All these results suggested that magnetic extraction method can be used for enrichment and quantification of RISP and 9OHR in biological samples.
SUPPLEMENTAL MATERIAL
Coming soon ....
REFERENCES (0)
CITATIONS (2)