Predictive Low-Glucose Suspend Reduces Hypoglycemia in Adults, Adolescents, and Children With Type 1 Diabetes in an At-Home Randomized Crossover Study: Results of the PROLOG Trial
Adult
Blood Glucose
Male
Cross-Over Studies
Adolescent
Blood Glucose Self-Monitoring
Monitoring, Ambulatory
Middle Aged
Hypoglycemia
3. Good health
Young Adult
03 medical and health sciences
Diabetes Mellitus, Type 1
Drug Delivery Systems
Insulin Infusion Systems
0302 clinical medicine
Humans
Hypoglycemic Agents
Insulin
Female
Child
Algorithms
Aged
DOI:
10.2337/dc18-0771
Publication Date:
2018-08-08T19:24:23Z
AUTHORS (18)
ABSTRACT
OBJECTIVE
This study evaluated a new insulin delivery system designed to reduce insulin delivery when trends in continuous glucose monitoring (CGM) glucose concentrations predict future hypoglycemia.
RESEARCH DESIGN AND METHODS
Individuals with type 1 diabetes (n = 103, age 6–72 years, mean HbA1c 7.3% [56 mmol/mol]) participated in a 6-week randomized crossover trial to evaluate the efficacy and safety of a Tandem Diabetes Care t:slim X2 pump with Basal-IQ integrated with a Dexcom G5 sensor and a predictive low-glucose suspend algorithm (PLGS) compared with sensor-augmented pump (SAP) therapy. The primary outcome was CGM-measured time <70 mg/dL.
RESULTS
Both study periods were completed by 99% of participants; median CGM usage exceeded 90% in both arms. Median time <70 mg/dL was reduced from 3.6% at baseline to 2.6% during the 3-week period in the PLGS arm compared with 3.2% in the SAP arm (difference [PLGS − SAP] = −0.8%, 95% CI −1.1 to −0.5, P < 0.001). The corresponding mean values were 4.4%, 3.1%, and 4.5%, respectively, represent-ing a 31% reduction in the time <70 mg/dL with PLGS. There was no increase in mean glucose concentration (159 vs. 159 mg/dL, P = 0.40) or percentage of time spent >180 mg/dL (32% vs. 33%, P = 0.12). One severe hypoglycemic event occurred in the SAP arm and none in the PLGS arm. Mean pump suspension time was 104 min/day.
CONCLUSIONS
The Tandem Diabetes Care Basal-IQ PLGS system significantly reduced hypoglycemia without rebound hyperglycemia, indicating that the system can benefit adults and youth with type 1 diabetes in improving glycemic control.