Validación de un método inmunológico para la determinación de vancomicina en líquido cefalorraquídeo
0303 health sciences
03 medical and health sciences
DOI:
10.31053/1853.0605.v76.n2.23550
Publication Date:
2019-06-19T20:31:12Z
AUTHORS (5)
ABSTRACT
Introduction: Vancomycin (VAN) is an antibiotic used to treat serious infections. Its use related adverse effects such as acute facial hyperemia, nephrotoxicity and ototoxicity. By having a very narrow therapeutic range, its monitoring necessary maximize efficiency minimize toxic effects. It estimated that concentration in CSF approximately 10% of the plasma level patients who receive intravenous treatment have meninges inflammation. Plasma concentrations VAN are not reliable indicator those present CSF. The aim this study was validate immunological method based on kinetic interaction microparticles solution (KIMS) for determination Materials Methods: KIMS validated evaluation For this, parameters linearity, precision, accuracy, limit detection, quantification, interference, selectivity specificity were determined. Results: linear range between 0 15 µg/mL, CV% obtained oscillated 0.7 2.5% range. LOD LOQ 0.4 µg/mL 1.4 respectively. equation line correlation methods HPLC-UV y = 0.9151x + 1.1695, R² 0.9453. Conclusion: demonstrated adequate sensitivity determine being useful tool complicated infections at CNS level.
SUPPLEMENTAL MATERIAL
Coming soon ....
REFERENCES (0)
CITATIONS (2)
EXTERNAL LINKS
PlumX Metrics
RECOMMENDATIONS
FAIR ASSESSMENT
Coming soon ....
JUPYTER LAB
Coming soon ....