An ultrasonic transducer focusing ultrasound into a thin waveguide by two elliptical reflectors
ultrasonic transducer
mode conversion
03 medical and health sciences
0302 clinical medicine
Physics
QC1-999
minimally invasive treatments
DOI:
10.35848/1882-0786/ad4d3c
Publication Date:
2024-05-17T22:40:38Z
AUTHORS (6)
ABSTRACT
Abstract A high-power ultrasonic transducer for minimally invasive treatments is proposed, capable of treating areas inaccessible by high-intensity focused ultrasound treatments. This employs two elliptical reflectors that efficiently focus waves into a thin waveguide utilizing mode conversion. Additionally, slit structure introduced to suppress wave diffraction, further enhancing the focusing capabilities transducer. Finite element analysis demonstrates proposed achieves an impressive energy density magnification factor 325, with efficiency 24.2 <mml:math xmlns:mml="http://www.w3.org/1998/Math/MathML" overflow="scroll"> <mml:mo>%</mml:mo> <mml:mo>.</mml:mo> </mml:math> 3.4 times higher than conventional transducer, Double Parabolic refLector wave-guided Ultrasonic