Nonrigid Patient Setup Errors in the Head-and-Neck Region

Radiotherapy Planning, Computer-Assisted Masks Radiotherapy Dosage Cone-Beam Computed Tomography Immobilization 03 medical and health sciences Radiation Protection 0302 clinical medicine Head and Neck Neoplasms Body Burden Humans Artifacts
DOI: 10.1007/s00066-007-1747-5 Publication Date: 2007-08-30T19:18:33Z
ABSTRACT
To investigate the magnitude and clinical relevance of relative motion/nonrigid setup errors in the head-and-neck (H&N) region.Eleven patients with tumors in the H&N region were immobilized in thermoplastic head masks. Patient positioning was verified using a kilovoltage cone-beam CT (kv CBCT) prior to 100 treatment fractions. Five different regions of interest (ROIs) were selected for automatic image registration of planning CT and verification CBCT: (1) the whole volume covering planning CT and CBCT, (2) the skull, (3) the mandible, (4) C1-C3, and (5) C4-C6. Differences were calculated describing relative motion between the ROIs.The 3-D patient setup error was 3.2 mm +/- 1.7 mm based on registration of the whole volume. No systematic relative motion (group mean errors <0.5 mm and <0.5 degrees ) between planning and treatment for any ROI was observed. Mobility was largest for the skull and the mandible relative to C4-C6 with 3-D displacements of 4.7 mm +/- 2.5 mm and 4.4 mm +/- 2.5 mm. Relative rotations were largest around the left-right axis (nodding) between C1-C3 and C4-C6 with maximum 11 degrees . No time trend of relative motion was observed. Margins for compensation of relative motion ranged between 5 mm and 10 mm.The simplification of the patient as a rigid body was shown to result in significant errors due to relative motion in the H&N region. Margins for compensation of relative motion exceeded margins for compensation of patient positioning errors.
SUPPLEMENTAL MATERIAL
Coming soon ....
REFERENCES (0)
CITATIONS (51)
EXTERNAL LINKS
PlumX Metrics
RECOMMENDATIONS
FAIR ASSESSMENT
Coming soon ....
JUPYTER LAB
Coming soon ....