Rapid spectrophotometric method to resolve ternary mixtures of Tartrazine, Quinoline Yellow and Patent Blue V in commercial products

04 agricultural and veterinary sciences 0405 other agricultural sciences
DOI: 10.1007/s002160051507 Publication Date: 2002-08-25T07:47:25Z
ABSTRACT
A very simple spectrophotometric method is described for resolving ternary mixtures of the food dyes Tartrazine (E-102), Quinoline Yellow (E-104) and Patent Blue V (E-131) by using the second derivative of the spectra with measurements at zero-crossing wavelengths. Calibration graphs are linear up to 20.0 mg/L of Tartrazine, up to 20.0 mg/L of Quinoline Yellow and up to 6.4 mg/L of Patent Blue V. Repeatability and reproducibility studies (with the Students’s and F tests) were achieved for two series of nine standards for each dye showing no significant differences at the 95% confidence level. Detection limits of 0.0526, 0.0164 and 0.0034 mg/L were obtained for Tartrazine, Quinoline Yellow and Patent Blue V, respectively. This method was used for determining synthetic mixtures of these colorants in different ratios and it was successfully applied to four commercial products without previous separation step. The results found in commercial products were compared with those obtained by an HPLC method and very similar values were found for both methods.
SUPPLEMENTAL MATERIAL
Coming soon ....
REFERENCES (0)
CITATIONS (21)