Non-invasive cardiac output evaluation during a maximal progressive exercise test, using a new impedance cardiograph device
Adult
Oxygen
03 medical and health sciences
Oxygen Consumption
0302 clinical medicine
Exercise Test
Humans
Reproducibility of Results
Cardiac Output
Cardiography, Impedance
DOI:
10.1007/s004210100458
Publication Date:
2003-02-12T23:21:55Z
AUTHORS (10)
ABSTRACT
One of the greatest challenges in exercise physiology is to develop a valid, reliable, non-invasive and affordable measurement of cardiac output (CO). The purpose of this study was to evaluate the reproducibility and accuracy of a new impedance cardiograph device, the Physio Flow, during a 1-min step incremental exercise test from rest to maximal peak effort. A group of 12 subjects was evaluated to determine the reproducibility of the method as follows: (1) each subject performed two comparable tests while their CO was measured by impedance cardiography using the new device (COImp1, COImp2), and (2) in a subgroup of 7 subjects CO was also determined by the direct Fick method (COFick) during the second test. The mean difference between the values obtained by impedance (i.e. COImp1-COImp2) was -0.009 l.min-1 (95% confidence interval: -4.2 l.min-1, 4.2 l.min-1), and CO ranged from 3.55 l.min-1 to 26.75 l.min-1 (n = 146). When expressed as a percentage, the difference (COImp1-COImp2) did not vary with increasing CO. The correlation coefficient between the values of COImp and COFick obtained during the second exercise test was r = 0.94 (P < 0.01, n = 50). The mean difference expressed as percentage was -2.78% (95% confidence interval: -27.44%, 21.78%). We conclude that COImp provides a clinically acceptable evaluation of CO in healthy subjects during an incremental exercise.
SUPPLEMENTAL MATERIAL
Coming soon ....
REFERENCES (0)
CITATIONS (203)
EXTERNAL LINKS
PlumX Metrics
RECOMMENDATIONS
FAIR ASSESSMENT
Coming soon ....
JUPYTER LAB
Coming soon ....