Accuracy of corrective osteotomy using a custom-designed device based on a novel computer simulation system
Models, Anatomic
Forearm
03 medical and health sciences
0302 clinical medicine
0206 medical engineering
Humans
Reproducibility of Results
Computer Simulation
Equipment Design
02 engineering and technology
Hand Deformities
Tomography, X-Ray Computed
Osteotomy
DOI:
10.1007/s00776-010-0020-4
Publication Date:
2011-02-25T07:33:30Z
AUTHORS (7)
ABSTRACT
We developed an original computer program that three-dimensionally simulates anatomical deformity correction of an extremity based on computed tomography (CT) data and an operative method that uses a custom-designed surgical device to realize the simulation. The purpose of this study was to investigate the feasibility and accuracy of this system using cadaver bones.Twenty dry forearm bones were used to investigate the accuracy of this system. The error in the setting location of the template was evaluated as the difference between the setting location in the computer simulation and the actual setting location. A mock corrective osteotomy using a custom-made template was performed and the results were compared with the preoperative simulation.All of the templates fitted snugly to the bone surface, and a mock osteotomy was performed as simulated. The errors in the setting location of the template on the bone surface were less than 1.0 mm and 1.0°. The accuracy of the simulation system as evaluated by a mock operation was also less than 1.0 mm and 1.0°.The system has sufficient feasibility and accuracy to realize anatomical and precise 3-D deformity correction of an extremity.
SUPPLEMENTAL MATERIAL
Coming soon ....
REFERENCES (30)
CITATIONS (49)
EXTERNAL LINKS
PlumX Metrics
RECOMMENDATIONS
FAIR ASSESSMENT
Coming soon ....
JUPYTER LAB
Coming soon ....