The Effect of Pulsed Electric Field on the Generation of Charge in Flow-Based Medical and Diagnostic Systems

Electric charge
DOI: 10.1007/s10527-018-9819-7 Publication Date: 2018-11-11T19:44:30Z
ABSTRACT
The aim of this study is to consider the effect of pulsed electric field on the process of charge generation caused by motion of water in flow-based systems widely used in medical proteomic and diagnostic devices. It has been demonstrated that the use of pulsed electric field leads to an increase in the value of charge generated during the motion of the liquid (~ 80%, T = 35°С). The observed effect should be taken into account in the development of novel highly sensitive flow-based diagnostic systems
SUPPLEMENTAL MATERIAL
Coming soon ....