Cardiac CT Angiography (CCTA) predicts left atrial appendage occluder device size and procedure outcome

Perimeter Caliber
DOI: 10.1007/s10554-016-1050-6 Publication Date: 2017-01-09T10:04:43Z
ABSTRACT
To investigate the role of cardiac CT angiography (CCTA) in predicting optimal left atrial appendage (LAA) occluder size and procedure outcome.Thirty-six patients underwent pre-procedural CCTA. CCTA and TEE LAA orifice diameters and perimeters were compared with the implanted device size. CCTA 3D configuration was correlated with procedure outcome. Watchman™ device (N = 18): diameters were 21 ± 4, 26 ± 5 and 25 ± 3 mm for TEE, CCTA and inserted device, respectively. Average perimeters were 61 ± 10, 74 ± 8 and 78 ± 11 mm for TEE, CCTA and inserted device, respectively. Better agreement with the device size was found for CCTA compared to TEE (Bland-Altman). ACP™ device (N = 15): diameters were 20 ± 5, 25 ± 4 and 23 ± 4 for TEE, CCTA and inserted device, respectively. Average perimeters were 58 ± 11, 72 ± 15 and 72 ± 13 mm for TEE, CCTA and inserted device, respectively. Excellent correlation and agreement with the device size was found for CCTA compared to TEE. CCTA perimeter >100 mm and "cactus" 3D configuration had a specificity of 96 and 81% respectively for procedure failure.CCTA LAA ostial perimeter predicted better the optimal occluder size as compared with the currently used LAA TEE diameter. Moreover, CCTA 3D data may help in predicting potential complications.
SUPPLEMENTAL MATERIAL
Coming soon ....
REFERENCES (28)
CITATIONS (27)