Calibration of gamma-ray spectrometers coupled to Compton suppression and fast pneumatic systems for the k0-standardized NAA method

01 natural sciences 0104 chemical sciences
DOI: 10.1007/s10967-008-1204-7 Publication Date: 2008-11-27T11:24:55Z
ABSTRACT
A new Compton suppression system (CSS) for the gamma-ray spectrometer portion of the neutron activation analysis (NAA) was set up at the RPI/ITN. The pneumatic transfer system, SIPRA, for short-lived nuclides and cyclic irradiations was improved. A full calibration procedure of the CSS and SIPRA systems was performed. Two certified reference materials, NIST-SRM-1572 (Citrus Leaves) and NIST-SRM-1633a (Coal Fly Ash) were analyzed using the calibration factors. The CSS was instrumental in lowering the detection limits of Cr, Fe, Hg, Rb, Sr, Th and Zn by reducing background and/or spectral interference considerably. The analytical results were evaluated by comparison to the NIST certified values with deviations ranging from 2% to 8% for the above mentioned elements, except Zn ranging from 10% to 15% for biological and environmental samples, respectively.
SUPPLEMENTAL MATERIAL
Coming soon ....
REFERENCES (8)
CITATIONS (8)