A novel minimal-contact biomotion method for long-term respiratory rate monitoring

Respiratory Rate Respiratory monitoring Sleep-Disordered Breathing
DOI: 10.1007/s11325-020-02067-4 Publication Date: 2020-04-15T11:03:08Z
ABSTRACT
In this study, we assessed the diagnostic accuracy of the device VitaLog (SWG Sportwerk GmbH & Co. KG, Dortmund, Germany) for estimation of respiratory rate (RR) variability.VitaLog is a minimal-contact biomotion device that is placed under the mattress topper. It senses respiratory effort and body movement using a piezoelectric sensor. Diagnostic accuracy was determined in 103 patients referred to our sleep laboratory for suspected sleep-disordered breathing (SDB). SDB was defined by AHI ≥ 15/h. Results provided by VitaLog were compared with nasal flow measurement obtained by polysomnography (PSG).Diagnostic accuracy of VitaLog was excellent. We obtained a correlation of r = 0.99 and a bias of 0.2 cycles per minute (cpm) between VitaLog and PSG-provided nasal flow. Detection RR variability worked nearly identically in patients with and without SDB.VitaLog is an appropriate method for determination of RR variability based on a minimal-contact biomotion sensor. This device is easy to handle, available at low cost, and suitable for long-term monitoring in the hospital or at home.
SUPPLEMENTAL MATERIAL
Coming soon ....
REFERENCES (22)
CITATIONS (8)
EXTERNAL LINKS
PlumX Metrics
RECOMMENDATIONS
FAIR ASSESSMENT
Coming soon ....
JUPYTER LAB
Coming soon ....