Quantification of stenotic mitral valve area with magnetic resonance imaging and comparison with Doppler ultrasound
Repeatability
Doppler ultrasound
Mitral valve stenosis
Cardiac magnetic resonance
DOI:
10.1016/j.jacc.2004.03.038
Publication Date:
2004-07-02T18:52:12Z
AUTHORS (9)
ABSTRACT
The purpose of this study was to evaluate the reliability of the pressure half-time (PHT) method for estimating mitral valve areas (MVAs) by velocity-encoded cardiovascular magnetic resonance (VE-CMR) and to compare the method with paired Doppler ultrasound.The pressure half-time Doppler echocardiography method is a practical technique for clinical evaluation of mitral stenosis. As CMR continues evolving as a routine clinical tool, its use for estimating MVA requires thorough evaluation.Seventeen patients with mitral stenosis underwent echocardiography and CMR. Using VE-CMR, MVA was estimated by PHT method. Additionally, peak E and peak A velocities were defined. Interobserver repeatability of VE-CMR was evaluated.By Doppler, MVAs ranged from 0.87 to 4.49 cm2; by CMR, 0.91 to 2.70 cm2, correlating well between modalities (r = 0.86). The correlation coefficient for peak E and peak A between modalities was 0.81 and 0.89, respectively. Velocity-encoded CMR data analysis provided robust, repeatable estimates of peak E, peak A, and MVA (r = 0.99, 0.99, and 0.96, respectively).Velocity-encoded cardiovascular magnetic resonance can be used routinely as a robust tool to quantify MVA via mitral flow velocity analysis with PHT method.
SUPPLEMENTAL MATERIAL
Coming soon ....
REFERENCES (23)
CITATIONS (62)
EXTERNAL LINKS
PlumX Metrics
RECOMMENDATIONS
FAIR ASSESSMENT
Coming soon ....
JUPYTER LAB
Coming soon ....