Pilot multi-site quality assurance study of batch-mode clinical-scale automated xenon-129 hyperpolarizers
Isotopes of xenon
DOI:
10.1016/j.jmr.2020.106755
Publication Date:
2020-05-29T15:35:11Z
AUTHORS (9)
ABSTRACT
We present a pilot quality assurance (QA) study of spin-exchange optical pumping (SEOP) performed on two nearly identical second-generation (GEN-2) automated batch-mode clinical-scale 129Xe hyperpolarizers, each utilizing a convective forced air oven, high-power (~170 W) continuous pump laser irradiation, and xenon-rich gas mixtures (~1.30 atm partial pressure). In one study, the repeatability of SEOP in a 1000 Torr Xe/900 Torr N2/100 Torr 4He (2000 Torr total pressure) gas mixture is evaluated over the course of ~700 gas loading cycles, with negligible decrease in performance during the first ~200 cycles, and with high 129Xe polarization levels (avg. %PXe = 71.7% with standard deviation σPXe = 1.5%), build-up rates (avg. γSEOP = 0.019 min-1 with standard deviation σγ = 0.003 min-1) and polarization lifetimes (avg. T1 = 90.5 min with standard deviation σT = 10.3 min) reported at moderate oven temperature of ~70 °C. Although the SEOP cell in this study exhibited a detectable performance decrease after 400 cycles, the cell continued to produce potentially useable HP 129Xe with %PXe = 42.3 ± 0.6% even after nearly 700 refill cycles. The possibility of "regenerating" "dormant" (i.e., not used for an extended period of time) SEOP cells using repeated temperature cycling methods to recover %PXe is also demonstrated. The quality and consistency of results show significant promise for translation to clinical-scale production of hyperpolarized 129Xe contrast agents for imaging and bio-sensing applications.
SUPPLEMENTAL MATERIAL
Coming soon ....
REFERENCES (61)
CITATIONS (9)
EXTERNAL LINKS
PlumX Metrics
RECOMMENDATIONS
FAIR ASSESSMENT
Coming soon ....
JUPYTER LAB
Coming soon ....