Analysis and correction of field fluctuations in fMRI data using field monitoring
2805 Cognitive Neuroscience
Adult
Echo-planar imaging
Magnetic field monitoring
Echo-Planar Imaging
Functional Neuroimaging
Magnetic Phenomena
fMRI
610 Medicine & health
530
Magnetic Resonance Imaging
Field fluctuations
170 Ethics
Physiological noise
Young Adult
03 medical and health sciences
0302 clinical medicine
2808 Neurology
Image Processing, Computer-Assisted
SFNR
Humans
10237 Institute of Biomedical Engineering
DOI:
10.1016/j.neuroimage.2017.01.014
Publication Date:
2017-01-09T00:45:16Z
AUTHORS (8)
ABSTRACT
This work investigates the role of magnetic field fluctuations as a confound in fMRI. In standard fMRI experiments with single-shot EPI acquisition at 3 Tesla the uniform and gradient components of the magnetic field were recorded with NMR field sensors. By principal component analysis it is found that differences of field evolution between the EPI readouts are explainable by few components relating to slow and within-shot field dynamics of hardware and physiological origin. The impact of fluctuating field components is studied by selective data correction and assessment of its influence on image fluctuation and SFNR. Physiological field fluctuations, attributed to breathing, were found to be small relative to those of hardware origin. The dominant confounds were hardware-related and attributable to magnet drift and thermal changes. In raw image time series, field fluctuation caused significant SFNR loss, reflected by a 67% gain upon correction. Large part of this correction can be accomplished by traditional image realignment, which addresses slow and spatially uniform field changes. With realignment, explicit field correction increased the SFNR on the order of 6%. In conclusion, field fluctuations are a relevant confound in fMRI and can be addressed effectively by retrospective data correction. Based on the physics involved it is anticipated that the advantage of full field correction increases with field strength, with non-Cartesian readouts, and upon phase-sensitive BOLD analysis.
SUPPLEMENTAL MATERIAL
Coming soon ....
REFERENCES (76)
CITATIONS (34)
EXTERNAL LINKS
PlumX Metrics
RECOMMENDATIONS
FAIR ASSESSMENT
Coming soon ....
JUPYTER LAB
Coming soon ....