Testing the Real-World Accuracy of the Dexcom G6 Pro CGM During the Insulin-Only Bionic Pancreas Pivotal Trial
Bionics
Blood Glucose
Diabetes Mellitus, Type 1
Insulin, Regular, Human
Blood Glucose Self-Monitoring
Humans
Insulin
Reproducibility of Results
Pancreas
3. Good health
DOI:
10.1089/dia.2023.0287
Publication Date:
2023-09-05T14:54:07Z
AUTHORS (6)
ABSTRACT
Continuous glucose monitors (CGMs) have transformed the way people with type 1 diabetes can self-monitor glucose levels. Past studies have evaluated the accuracy of CGMs in clinic-based studies, but few have analyzed their accuracy in real-world settings. The Insulin-Only Bionic Pancreas Trial provided the opportunity to assess real-world accuracy of the blinded Dexcom G6 Pro sensor over the first 48-60 h of wear using a blood glucose meter (BGM) as a comparator for 1073 CGM-BGM pairs across 53 participants. The mean absolute relative difference (MARD) was 11.0% over a median period of 50 h (range 47-79 h). The MARD was 13.6% in the first 12 h, 10.5% in hours 12-24, and 10.1% after the first 24 h. These results are comparable with accuracy shown previously with laboratory-based measurements and provide real-world evidence of Dexcom G6 Pro accuracy, which improved after the first 12 h and then remained stable thereafter. Clinical Trial Registry: clinicaltrials.gov; NCT04200313.