MEASUREMENT OF NEUTRON AMBIENT DOSE EQUIVALENT IN PROTON RADIOTHERAPY WITH LINE-SCANNING AND WOBBLING MODE TREATMENT SYSTEM
Isocenter
Pencil-beam scanning
Equivalent dose
Dose profile
DOI:
10.1093/rpd/ncx056
Publication Date:
2017-04-07T07:08:00Z
AUTHORS (9)
ABSTRACT
The primary objective of this study was to measure secondary neutron dose during proton therapy using a detector that covers the entire energy range produced in therapy. We analyzed and compared treatment with passive scattering line scanning. ambient equivalents were measured 190 MeV wobbling line-scanning beam. center plastic water phantom (30 × 30 60 cm3) placed at isocenter. A Wide-Energy Neutron Detection Instrument (WENDI-2) located 1m from isocenter four different angles (0°, 45°, 90° 135°). Both modes multipurpose pencil beam scanning dedicated nozzles used obtain spread-out Bragg peak 10-cm-width for measurements. equivalent H*(10) value normalized by therapeutic For mode mode, highest values 1.972 0.099 mSv/Gy, respectively. successfully six positions generated WENDI-2. These reference data could be reduction methods other analysis advanced near future.