Wide field-of-view Talbot grid-based microscopy for multicolor fluorescence imaging
610
Breast Neoplasms
Equipment Design
Image Enhancement
530
01 natural sciences
Equipment Failure Analysis
Microscopy, Fluorescence, Multiphoton
Cell Line, Tumor
0103 physical sciences
Humans
Lighting
DOI:
10.1364/oe.21.014555
Publication Date:
2013-06-11T17:04:49Z
AUTHORS (5)
ABSTRACT
The capability to perform multicolor, wide field-of-view (FOV) fluorescence microscopy imaging is important in screening and pathology applications. We developed a microscopic slide-imaging system that can achieve multicolor, wide FOV, fluorescence imaging based on the Talbot effect. In this system, a light-spot grid generated by the Talbot effect illuminates the sample. By tilting the excitation beam, the Talbot-focused spot scans across the sample. The images are reconstructed by collecting the fluorescence emissions that correspond to each focused spot with a relay optics arrangement. The prototype system achieved an FOV of 12 × 10 mm(2) at an acquisition time as fast as 23 s for one fluorescence channel. The resolution is fundamentally limited by spot size, with a demonstrated full-width at half-maximum spot diameter of 1.2 μm. The prototype was used to nimage green fluorescent beads, double-stained human breast cancer SK-BR-3 cells, Giardia lamblia cysts, and the Cryptosporidium parvum oocysts. This imaging method is scalable and simple for implementation of high-speed wide FOV fluorescence microscopy.