Compact quantum cascade laser based quartz-enhanced photoacoustic spectroscopy sensor system for detection of carbon disulfide
Quantum cascade laser
Photoacoustic Spectroscopy
Trace gas
Carbon disulfide
Parts-per notation
Continuous wave
DOI:
10.1364/oe.24.006559
Publication Date:
2016-03-16T15:07:15Z
AUTHORS (3)
ABSTRACT
A compact gas sensor system based on quartz-enhanced photoacoustic spectroscopy (QEPAS) employing a continuous wave (CW) distributed feedback quantum cascade laser (DFB-QCL) operating at 4.59 µm was developed for detection of carbon disulfide (CS2) in air trace concentration. The influence water vapor monitored QEPAS signal investigated to enable compensation this dependence by independent moisture sensing. 1 σ limit 28 parts per billion volume (ppbv) s lock-in amplifier time constant obtained the CS2 line centered 2178.69 cm-1 when sample moisturized with 2.3 vol% H2O. work reports suitability monitoring high selectivity and sensitivity, as well low requirements fast response applications such workplace process industry.