A Mail Audit Independent Peer Review System for Dosimetry Verification of a Small Animal Irradiator
Phantoms, Imaging
Peer Review
X-Rays
Radiobiology
Radiation Dosage
Mice
03 medical and health sciences
0302 clinical medicine
Calibration
Animals
Postal Service
Radiometry
DOI:
10.1667/rr15220.1
Publication Date:
2020-02-18T16:32:13Z
AUTHORS (13)
ABSTRACT
Dedicated precision orthovoltage small animal irradiators have become widely available in the past decade and are commonly used for radiation biology research. However, there is a lack of dosimetric standardization among these irradiators, which affects reproducibility radiation-based studies. The purpose this study was to develop mail-based, independent peer review system verify dose delivery institutions using X-RAD 225Cx (Precision X-Ray, North Branford, CT). A robust, user-friendly mouse phantom constructed from high-impact polystyrene designed with dimensions similar those typical laboratory mouse. accommodates three thermoluminescent dosimeters (TLDs) measure dose. commissioned irradiator anterior-posterior posterior-anterior beams 225 kVp then mailed test feasibility audit service. energy correction factor TLDs derived validate delivered particular irradiation system. This indicated that were able deliver within ±10% target developed mail verification dosimetry can be expanded characterize other commercially thereby enhancing studies employing irradiators.
SUPPLEMENTAL MATERIAL
Coming soon ....
REFERENCES (22)
CITATIONS (6)
EXTERNAL LINKS
PlumX Metrics
RECOMMENDATIONS
FAIR ASSESSMENT
Coming soon ....
JUPYTER LAB
Coming soon ....