Broadband Time-Resolved Absorption and Dispersion Spectroscopy of Methane and Ethane in A Plasma Using a Mid-Infrared Dual-Comb Spectrometer
Spectral resolution
Temporal resolution
DOI:
10.20944/preprints202009.0397.v1
Publication Date:
2020-09-17T13:59:11Z
AUTHORS (6)
ABSTRACT
Conventional mechanical Fourier Transform Spectrometers (FTS) are able to simultaneously measure absorption and dispersion spectra of gas-phase samples. However, they usually need very long measurement times achieve time-resolved with a good spectral temporal resolution. Here, we present mid-infrared dual-comb-based FTS in an asymmetric configuration, providing broadband resolution 5 GHz, 20 μs, total time few minutes. We used the dual-comb spectrometer monitor reaction dynamics methane ethane electrical plasma discharge. observed ethane/methane formation as recombination hydrocarbon radicals discharge various static dynamic conditions. The results demonstrate new analytical approach for measuring fast molecular changes monitoring chemical reactions, which can be interesting kinetic research particularly combustion analysis community.
SUPPLEMENTAL MATERIAL
Coming soon ....
REFERENCES (0)
CITATIONS (2)
EXTERNAL LINKS
PlumX Metrics
RECOMMENDATIONS
FAIR ASSESSMENT
Coming soon ....
JUPYTER LAB