Control of Blood Glucose in Experimental Diabetes by Means of a Totally Implantable Insulin Infusion Device
Blood Glucose
Drug Implants
03 medical and health sciences
Dogs
Pancreatectomy
0302 clinical medicine
Animals
Insulin
Artificial Organs
Glucose Tolerance Test
Diabetes Mellitus, Experimental
3. Good health
DOI:
10.2337/diab.28.7.634
Publication Date:
2013-09-19T17:13:09Z
AUTHORS (7)
ABSTRACT
Near-normal glucose tolerance tests in diabetic dogs were obtained during basal rate insulin infusions in restrained animals by use of extracorporeal infusion pumps and in conscious, unrestrained animals by means of implanted infusion pumps. Even better regulation of blood glucose in diabetic animals was obtained by the addition of predetermined pulses of insulin at higher flow rates than the basal flow rate, accomplished by use of a transcutaneously activated valve mechanism attached to the implanted infusion pump. We conclude that near-normal blood glucose concentrations can be maintained throughout the day in the dog by these means and that similar approaches, using implantable infusion pumps, in man may lead to better long-term control of diabetes than is currently available.