The influence of acquisition mode on the dosimetric performance of an amorphous silicon electronic portal imaging device
integrated acquisition mode
Medical physics. Medical radiology. Nuclear medicine
03 medical and health sciences
0302 clinical medicine
R895-920
Technical Note
Cine acquisition mode
electronic portal imaging device dosimetry
DOI:
10.4103/jmp.jmp_98_16
Publication Date:
2017-06-16T09:12:58Z
AUTHORS (5)
ABSTRACT
This study investigates the impact of cine acquisition mode on the dosimetric characteristics of a Varian aS500 amorphous silicon electronic portal imaging device (a-Si EPID).The performance of an a-Si EPID operated in cine mode was assessed and compared to its performance when operated in an integrated mode and dose measurements using an ionization chamber. This study was conducted at different photon energies and the EPID performance was assessed as function of the delivered dose, dose rate, multileaf collimator speed, field size, phantom thickness, and intensity-modulated radiation therapy fields.The worst nonlinearity was observed at low monitor unit (MU) settings < 100 MU with the highest dose per frame. The nonlinearity of response at a low MU setting was attributed due to the loss of four cine images during each delivery. The EPID response with changing dose rate for 10 MU delivered had similar results to its performance in an integrated mode and ionization chamber. Despite the nonlinearity of response with low MU delivered, EPID performance operated in cine and integrated acquisition modes had comparable responses within 2%.For EPID dosimetry application using cine mode, this study recommends the calibration of the EPID images to be undertaken at a large MU. There were no additional corrections that were required when the EPID operated in cine acquisition mode as compared to calibration in integrated mode.
SUPPLEMENTAL MATERIAL
Coming soon ....
REFERENCES (18)
CITATIONS (4)
EXTERNAL LINKS
PlumX Metrics
RECOMMENDATIONS
FAIR ASSESSMENT
Coming soon ....
JUPYTER LAB
Coming soon ....