Main text The report summarises the results of CCT-K8 key comparison in field humidity. main target this was to obtain degree equivalence (DoE) between realizations local scales dew point temperature humid gas (air), range from 30 °C 95 °C, among participating national metrology institutes. gives details scheme, procedure and analysis comparison. overall display a very good agreement nine laboratories. show excellent realized with no detected outliers. performance travelling standards can be...
Main text The report summarises the results of EURAMET.T-K8 key comparison (EURAMET Project No. P 717) in field humidity. covers dew-point temperature range from +30 °C up to + 95 dew point temperature. gives details scheme, procedure and analysis comparison. overall display a very good agreement 16 participating laboratories. two hygrometers used as travelling standard were sufficiently stable. linkage corresponding CCT-K8 will be calculated when final for both comparisons are available....